TDR System - Technical Details


Number of Test-Stations 1, including Friability Drum with opening
Loading and
Emptying of the Drums
The Friability Drum/s is loaded and emptied without removing and opening of the Drum. Due to the special Design the Friability Drum is loaded via an opening and on completion of the preset Run-Time the Samples are automatically discharged through the opening by turning the Drum backwards.
Turning Speed Adjustable from 20 to 100 rpm, accuracy +/- 1 rpm
Keypad Numeric Membrane Keypad with Function Keys
Display 4-line illuminated LC Display for control and display of RPM during and display of Results on completion of the Test-Run
Pre-Selection of Test-Run Time In Time: 1 sec. to 9 hrs., 59min., 59 sec.
In Revolutions: 1 to 99.999
Operation The operation is completely menu guided
Results When a Balance is connected to the unit the TDR displays Sample Weight before and after the Test-Run as well as calculates the lost. Results and Documentation of RPM can be hard-copied when a suitable Printer is connected to the standard parallel Printer Port.
Interface Serial RS 232 Interface for connection of a Balance (see Options)

Parallel Printer Port for hard-copy of Results and RPM documentation
10° angle for bigger Samples The TDR is equipped with adjustable feet for the angle adjustment of the Drum in accordance with USP/EP requirements